UPPER LEVEL ASSEMBLY: Disassembly Procedure
Revision A
DASH 2000 Patient Monitor
7-3
2000 412-001
Disassembly Procedure
Preparations before
Opening the monitor
Before any service interventions, turn off the device and disconnect the
device from power line.
Opening Unit
To open the unit, do the following.
1.
Lay the monitor with the front panel face down on a clean, level
surface (ESD pad) which is placed on a soft material to avoid
scratches on the front panel.
2.
Undo the 6 screws on the rear to remove the front panel frame and
open the device.
3.
Fold the front panel frame down and disconnect the keypad
connector.
Display
To remove the display, do the following.
1.
Open the unit as described in
Opening Unit
.
2.
Remove the 2-pole converter plug and the flat cable connector for the
display from the PCB Mainboard.
3.
Undo the screws which hold the converter.
4.
Undo the 4 screws which hold the display and remove it.
DAS PCB
To remove the DAS PCB, do the following.
1.
Open the unit as described in
Opening Unit
.
2.
Remove the 2-pole converter plug and the flat cable connector for the
display from the PCB Mainboard.
3.
Tilt the display frame with the DAS board towards the front.
4.
Disconnect the DAS flat cable connector.
5.
Remove NIBP tubing from DAS PCB.
6.
Undo the 4 screws which hold the DAS PCB and remove it.
DAS Input Assembly
To remove the DAS Input Assembly, do the following.
1.
Open the unit as described in
Opening Unit
.
2.
Remove the 2-pole converter plug and the flat cable connector for the
display from the PCB Mainboard.
3.
Tilt the display frame with the DAS board towards the front.
4.
Loosen the DAS input assembly and remove it
5.
Remove NIBP tubing from DAS input assembly.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......