CONFIGURATION: Loading Software
Revision A
DASH 2000 Patient Monitor
6-3
2000 412-001
Loading Software
Intended Use
This part of the section is for the purpose of loading manufacturer
software into the monitor initially, reloading software when the
possibility of corrupted software exists, or updating software in the event
of a release of a new software revision.
Software Loading/
Updating Methods
The process of loading or updating software in the monitor is described in
this part of the section. Manufacturer software can be loaded into the
monitor using these methods:
From Diskette
¥
The monitor is connected directly to a personal computer (PC) or PC
laptop. The Update Program is run off of the update diskettes and
the software is downloaded to the monitor via serial communication.
Over the Network
Over the Network
¥
For the monitor connected to a patient monitoring network, the
software is loaded from the update diskettes onto a Centralscope
central station or a Clinical Information Center (CIC). The central
station or CIC then acts as a network file server and software is
downloaded to the monitor over the network.
NOTE
Each method of downloading software to the monitor is
distinctly different. Completely read all of this part of the
section prior to any attempt to load or update software.
This is particularly important if this is a first attempt to
load or update software in the monitor.
Software
Compatibility
Write down or print out software code part numbers from the
SOFTWARE SUMMARY window for each monitor in the system. To
print the SOFTWARE SUMMARY table from each monitor, use the Trim
Knob to select the following menu option items from the monitor main
menu display:
MONITOR SETUP,
SOFTWARE SUMMARY, then
press the GRAPH GO/STOP key.
To print the software revision of the writer, use the Trim Knob to scroll to
and select the following menu option items from the monitor main menu
display:
MONITOR SETUP (1),
SERVICE MODE (2),
enter password (3)
(The password is the current date shown in the upper left
corner of the display. The expected format is: DDMM.)
CALIBRATE AND TEST (4),
GRAPH TEST PATTERN (5), then
select START (6)
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......