Using an SD Card
Restoring Files from the SD Card
After files have been copied to an SD card, you can use the
Restore All
function to
copy files from the SD card to internal storage.
NOTE:
If
CT Data Guard
is enabled and its
Prevent deleting of untransmitted records
field is set to
Yes
, you will receive the following warning if any of the records in
internal storage have not been transmitted:
The operation deletes existing records. This cannot be done since there are
untransmitted records!
If you receive this message, make sure that all records have been transmitted and try
Restore All
again.
1.
Insert an SD card with ECGs into the system's card slot.
Refer to
“Inserting and Ejecting the SD Card” on page 100
2.
Select
File Manager
to open the list of stored ECGs.
3.
Select
Restore All
.
The following message is displayed:
Existing records (if any) in memory will be deleted. Continue?
4.
Do one of the following:
• To cancel the copy, select
No
.
• To continue copying the files, select
Yes
.
The following message is displayed:
Deleting existing records (if any) in memory.... Please wait.
An indicator shows the progress/completion of the copying procedure.
Saving Files in XML Format
In addition to transmitting ECG records as XML files to a PC (see
XML Format” on page 86
), the MAC system can save ECG records as XML files on an
SD card.
NOTE:
Only ECGs recorded at 500 Hz or 500 Hz DVS can be saved in XML format. ECGs
recorded at 250 Hz cannot be saved in XML format.
1.
Insert an SD card into the system's card slot.
Refer to
“Inserting and Ejecting the SD Card” on page 100
2.
Select
File Manager
to open the list of stored ECGs.
3.
Highlight the record(s) to export as XML.
4.
Select
Save XML
.
The ECGs are converted to XML and saved to the SD card.
102
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......