Arrhythmia Mode Recording
Arrhythmia Recording Options
Arrhythmia Options- first row
Option
Description
Start Recording
Starts the arrhythmia reading.
If you did not fill out the Patient Data to select a patient, you
receive the following message:
No Patient Selected. Do you
want to continue without patient data?
1.
Select the
No
tab to continue.
The
Enter Patient Data
window opens.
2.
Enter the information on each page and select
Save
.
3.
Select
Start Recording
.
25 mm/s
NOTE:
The initial
measurement
displayed is
set in
System
Configuration
>
Arrhythmia Setup
.
Changes the speed of the waveform on the display and
printout. Changing the speed also changes the speed at which
the wiper bar moves across the display.
Measurement is in millimeters per second (mm/s) and includes
the following options:
• 25 mm/s
• 50 mm/s
• 12.5 mm/s - 5 mm/s
• 12.5 mm/s
When the option includes two speeds (12.5 mm/s - 5 mm/s),
the first speed is for the display and the second speed is for
the printout.
Changing the measurement here does not change the
measurement set in
System Configuration
.
5 mm/mV
NOTE:
The initial
measurement
displayed is
set in
System
Configuration
>
Arrhythmia Setup
.
Changes the magnitude of the ECG signal on the display or in
the report. Measurement is in millimeters per millivolt (mm/mV)
and includes the following options:
• 5 mm/mV
• 10 mm/mV
• 20 mm/mV
• 40 mm/mV
• 2.5 mm/mV
• Automatic
The larger the selected measurement, the larger the waveform.
Only the representation of the waveform changes; signal
strength is not affected.
NOTE:
If
Automatic
is selected, the system calculates the best
gain based on the peak-to-peak amplitudes of all displayed
leads and the selected display format.
Changing the measurement here does not change the
measurement set in
System Configuration
.
84
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......