Preparing the Patient
NEHB Placement
To acquire a NEHB ECG, use the standard 12–lead electrode placement and items 1
and 2 as shown in the following diagram.
NEHB Electrode Placement
AHA Label
IEC Label
Description
1
A1 orange
Nst white
Attachment point of the second rib to the right sternal
edge
2
A2 orange
Nax white
Fifth intercostal space on the left posterior axillary line
(Same position as V7 or C7)
3
V4 blue
Nap white
Mid-clavicular line in the fifth intercostal space (Same
position as C4)
54
MAC™ 2000
2053535-002K