Setting Up the Equipment
This system is compatible with MUSE v7.1.1, v8.0.1, and v9.0.0 and with CardioSoft/CS
v6.51, v6.61, v6.71 and v6.73.
Inserting the Paper
Before you can print ECG reports, complete the following steps:
1.
Make sure the system is set up for the correct paper size.
This device can print on the following papers: A4, standard letter (8.5 x 11
inches), or modified letter (8.433 x 11 inches).
For information on adjusting the printer for the paper size, see
Tray for Paper Size” on page 177
2.
Insert the appropriately sized paper.
Turning on the System
1.
Press the power button to turn on the system.
2.
Verify the system welcome screen is displayed with no errors.
NOTE:
If you encounter any problems powering on the system, see
Not Power Up” on page 184
for further troubleshooting instructions.
Configuring the Device
When the device is ready for operation, configure the system settings using the
information in
“System Configuration” on page 113
If you are applying the same settings to multiple devices at the site, export the settings
to an SD card and use that card to import the settings to other systems.
Testing the Device
After you have set up and configured the device, test the device completely before
using it with patients. Use the following test scenarios:
• Conducting and printing a resting ECG
See
“Recording a Resting ECG” on page 69
for instructions.
• Conducting and printing an arrhythmia ECG
See
“Arrhythmia Mode Recording” on page 83
for instructions.
• Conducting and printing a stress ECG.
See
“Stress Testing” on page 95
for instructions.
• Saving, importing, printing, deleting, transmitting, and exporting records.
See
“Managing Internal Storage” on page 105
for instructions.
50
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......