Product Overview
ECG Data Acquisition
ECG Data Acquisition provides the following:
• Samples with a minimum 500 Hz or 1000 Hz to the ECG processing algorithms
and the application software
• Pace enhancement enable/disable through the user interface
• QRS detection and heart rate calculation
• Lead sequences with 6 or 12 leads, where each lead is an element of the set (I, II, III,
aVR, aVL, aVF, D, A, J, V1…V6) with an optional –aVR
ECG Data Acquisition supports the following:
• Default high pass filter (0.04 Hz), if ADS is on high pass filter (0.56 Hz)
• Selectable low pass filter (20, 40, 100, 150 Hz)
• Selectable mains filter (50 Hz, 60 Hz)
• Anti Drift System (ADS): Baseline shift correction with finite impulse response high
pass filter enable/disable through the user interface in Resting ECG, Stress, and
RR analysis modes
The following are selectable data formats for external ECG storage:
• DCAR XML, 500 Hz uncompressed
• DCAR XML, 1000 Hz uncompressed
• Hilltop, 500 Hz DVS
External Storage
This system supports a Secure Digital High Capacity (SDHC) card with 4 GB capacity as
external storage.
2053535-002K
MAC™ 2000
39
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......