Maintenance
7.
Remove the AC mains power and turn on the device.
8.
Allow the battery to discharge until the device shuts down.
9.
Reconnect the AC mains power to the device and leave the device turned off.
10.
Allow the battery to fully charge.
When the
Battery LED
indicator stops flashing and shines steadily, the battery
is fully charged and the conditioning cycle is complete.
Battery Status Indicator
The battery status indicator is located on the top of the screen. For the exact location,
see the screen in
“Recording a Resting ECG” on page 69
.
The following diagram and table describe the battery status.
Battery Status
Item
Description
1
The battery is fully charged and above 75%.
2
The battery charge is above 50%.
3
The battery charge is above 25%.
4
The battery charge is below 25%.
This status is also used when the battery charge is unknown.
Supplies and Accessories
For a list of available supplies and accessories, refer to the supplies and accessories
reference guide for this device.
2053535-002K
MAC™ 2000
181
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......