Stress Testing
Stress Text Option Keys—Second Row (cont'd.)
Option
Description
Select Protocol
Selects a predefined set of test criteria. For more information, see
“Editing Stress Protocols” on page 137
.
Report Format
Selects the components and episodes to include in the report.
Allows you to override the defaults set on the
Stress ECG Setup
window. For more information, see
“Stress ECG Settings” on page
.
Target HR
Enter the maximum heart rate calculated for the patient based on
weight, gender, age, and condition. The ECG cart system monitors
the heart rate against this target.
More
Toggles between the first, second, and third row of options.
Stress Text Option Keys—Third Row
Option
Description
Main Menu
Exits the
Stress ECG
function and returns to the
Main Menu
.
More
Toggles between the first, second, and third row of options.
Conducting Stress Tests
There are two basic processes for conducting a stress test:
• Conducting a stress test with a treadmill or ergometer
• Conducting a stress test with a Master’s Step device
Each process is described in this section. For information on the
Stress Mode
interface,
see
“Stress Mode Interface” on page 96
Conducting a Stress Test with a Treadmill or Ergometer
Use the following instructions to conduct a stress test with a treadmill or ergometer.
The process is essentially identical for all devices with only minor differences between
supported equipment, unsupported equipment, and ergometers with remote start.
Deviations for specific accessories are noted where appropriate.
WARNING:
PATIENT INJURY — When on a moving treadmill, a patient could fall and sustain
an injury.
To minimize the possibility of a falling caused by the belt’s sudden movement,
have the patient step onto the belt only after it begins moving.
When conducting stress tests on a supported treadmill, press the
Stop TM
button
twice to immediately stop the belt in the case of an emergency (for example, if
the patient stumbles or falls while the belt is moving).
1.
Prepare the patient as described in
“Preparing the Patient” on page 51
2.
On the
Main Menu
press the
Stress ECG
option.
The
Enter Patient Data
window opens.
100
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......