5.4.2 Impedance respiration
The messages below appear in the Resp parameter window (PW), ImpResp waveform field (WF)
or the message field (MF).
5.4.3 Invasive Pressure
The messages below appear in the InvBP parameter window (PW), InvBP waveform field (WF) or
the message field (MF).
Message
Location
Possible causes
Possible solutions
Measurement off
WF, PW
ECG cable is disconnected.
Module is disconnected.
Impedance respiration PW or
WF is not configured on the
screen.
One or more leads required for
impedance measurement are
disconnected.
Impedance Respiration
Measurement is “Off”.
Connect the ECG cable.
Connect the module.
Configure either the ImpResp
waveform or Resp parameter
window on the monitor screen.
Connect any disconnected lead
wires and/or electrodes.
Set Impedance Respiration
Measurement “On” from the
Setup tab in the Impedance
respiration menu.
RA/R lead off
LA/L lead off
LL/F lead off
WF, PW
One of the following lead wires
or electrodes is disconnected:
RA, LA, or LL.
Check lead wire and electrode
connections.
Small resp curve
PW
Signal amplitude < 0,4 Ohm
Check lead wire and electrode
connections.
Apnea deactivated
PW
Apnea alarm start-up conditions
are not reached.
Apnea alarm detection is
activated after the 5 breaths are
detected.
Message
Location
Possible causes
Possible solutions
No Px transducer
NOTE: The 'x' in Px refers
to the InvBP channel
number.
MF
Pressure transducer and/or
transducer adapter cable is/are
disconnected from the module.
Connect pressure transducer
and/or transducer adapter
cable to the module.
Not zeroed
IPs not zeroed
PW, WF
MF
There is at least one invasive
pressure channel, which is not
zeroed.
Zero all invasive pressure
channels.
Px over range
>320mmHg' or '>43 kPa
NOTE: The 'x' in Px refers
to the InvBP channel
number.
MF
PW
Measured pressure exceeds 320
mmHg / 43 kPa upper
specification limit.
Transducer is not zeroed
correctly.
The measurement range is
between -40 mmHg and 320
mmHg.
Zero the invasive pressure
channel x.
Summary of Contents for E-PSMP
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes...
Page 6: ...Module Frames and Modules 1 ii 2062973 004...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Page 199: ...6 Masimo Module E MASIMO Technical Manual...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004...
Page 236: ...Module Frames and Modules 7 12 2062973 004...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Page 381: ...10 Single width Airway Module E miniC Technical Manual...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Page 561: ...16 Patient Data Module v2 Technical Manual...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004...
Page 621: ......