Configuration
67
2062973-003
7
Configuration
The configuration of the patient monitor consists of platform configuration and clinical
configuration.
This chapter describes how to perform the platform configuration tasks needed to take the
patient monitor into use for the first time and the configuration tasks needed for administration
and maintenance thereafter.
For information on how to perform the clinical configuration, including care unit settings and
user profiles, refer to the patient monitor’s user’s manual.
7.1 Adjusting display
7.1.1 Adjusting the brightness of the integrated primary display
If needed, adjust the display brightness from the
Monitor Setup > Brightness
menu. Refer to
the patient monitor’s user’s manual for more information.
7.1.2 Calibrating a touchscreen
NOTE: The touchscreen function of the integrated primary screen has been calibrated in the
factory. Re-calibration is typically needed only after replacing the uDOM or the touchscreen
sensor.
Calibrate the connected touchscreens, one by one, as follows:
1.
Select
Monitor Setup > Service Calibrations.
2.
In the
Enter Password
dialog box, type the username and password and press
Enter
.
Username:
biomed
Password:
Change
<space>
Me
NOTE: Username and password are case sensitive.
NOTE: The factory default password for the username “biomed” is
“
Change Me”.
3.
In the
Service / Calibrations
menu, select
Touch Screen
to calibrate the integrated
primary display or
External Touch Screen
to calibrate the connected secondary display.
4.
The Touch calibration screen opens.
5.
Touch the calibration mark or cross hair (+) in each corner of the screen, according to the
instructions shown on the screen.
7.1.3 Adjusting optional secondary display
If needed, adjust the picture on the display using the Auto Adjustment feature in the display’s
OSD menu. Refer to the display's user manual for details.
7.2 Configuring the network
7.2.1 Configuring hostname
1.
Log in to Webmin
2.
Select
Configuration >
Network > Hostname.
The
Hostname Configuration
window opens.
The current hostname is shown in the
Current Value
column.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......