Hardware installation
61
2062973-003
2.
Connect the Module Bus Adapter for PSM to the PSM connector in the patient monitor.
3.
Connect the cable from the Pole Mount for PSM to the Module Bus Adapter for PSM.
Installing a PDM module to a mounting solution:
Ensure that the selected PDM mount is properly installed according to the installation
instructions included with the mounting hardware.
1.
Connect the PDM module to installed mounting hardware as instructed in the
accompanying installation instructions.
2.
Connect one end of the ePort cable to the PDM module ePort connector.
3.
Connect the other end of the ePort cable to the ePort connector in the rear panel of the
patient monitor.
WARNING
PHYSICAL INJURY — Do not install the PDM above a patient. Make sure the
battery is completely inserted and the battery door is completely closed.
Falling batteries could seriously or fatally injure neonatal or other
vulnerable patients.
WARNING
PHYSICAL INJURY — Do not install the PDM above a patient. Leaks from the
battery cells can occur under extreme conditions. The liquid is caustic to the
eyes and skin. If the liquid comes in contact with eyes or skin, flush with
clean water and seek medical attention.
Installing E-modules to the patient monitor
1.
Connect a module by aligning it with the insertion guide inside the module slot.
2.
Push the module into the frame until it locks.
6.5 Connecting to the mains power
WARNING
Use only AC power cords recommended or manufactured by GE.
WARNING
EXCESSIVE LEAKAGE CURRENT - To avoid summation of leakage currents
when interfacing the device with other equipment, the devices may only be
interconnected with each other or to parts of the system when it has been
determined by qualified biomedical personnel that there is no danger to the
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......