CARESCAPE Monitor B450
56
2062973-003
−
Three LEDs illuminated: 50% – 74.9% of full-charge capacity.
−
Two LEDs illuminated: 25% – 49.9% of full-charge capacity.
−
One LED illuminated: 10% – 24.9% of full-charge capacity.
−
One LED flashing: < 10% of full-charge capacity remaining.
NOTE: Before taking the patient monitor into use for the first time, the batteries should be fully
charged. Keep the PDM connected to the patient monitor and the patient monitor connected to
the AC mains until the batteries are fully charged.
6.1.2 Installing the battery into the patient monitor
WARNING
The patient monitor must always be used with a battery inserted. This will
ensure the functioning of the monitor and prevent data loss during possible
supply mains interruptions.
1.
Open the battery cover by pressing the battery cover release latch down and pulling the
battery door open.
2.
Insert the batteries, one at time, with the test indicator facing to front and the connector
end first all the way into the battery slot.
3.
Close the battery door carefully.
6.1.3 Installing the battery into the PDM module
1.
Open the battery door by gently pulling on the battery door pull tab.
2.
Pull the battery tray out of the PDM using the battery tray strap.
3.
Insert the battery with the
TEST
button facing up and the arrow pointing into the PDM.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......