Disassembly and reassembly
195
2062973-003
12.2.11 Detaching the front unit assembly from the mid-frame assembly
Disassemble first:
a.
12.2.6. Detaching the front unit and mid-frame assembly from the rear unit assembly
1. Remove the screw cover plate at the bottom
of the front panel.
2. Remove the two screws (T10) holding the front
unit assembly to the mid-frame assembly.
Keep the front unit and mid-frame assembly on
the table the display facing down. Protect the
touchscreen sensor from any scratches and dust.
3. Detach user interface and touchscreen cables
from the DC/DC board.
4. Release the four snaps - two on each side -
that mount the front unit assembly to the
mid-frame assembly.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......