Configuration
99
2062973-003
activation is completed and the parameter module has restarted. This may take up to 15
minutes. Do not shut down the patient monitor or disconnect the parameter modules.
The parameter module restarts automatically after the software activation is complete.
4.
After the parameter module restarts, verify that the software activation is successful and
the parameter module runs the activated software.
7.19.3 Canceling pending host software activation
You can cancel a pending host software activation at any time while the message
Software
activation after next discharge/after next case end
is shown on the display.
1.
Log in to Webmin.
2.
Select
Configuration
>
Software Management.
The
Software Management
window opens.
3.
Select
Host Software
from the
Software List
and select
Next
.
4.
In the
Host Software
window, below
Status
, verify that the
Software activation pending
is
Yes
.
5.
Select
Cancel
.
The pending software activation is cancelled and the patient monitor continues running the
current software version.
7.19.4 Erasing an inactive software version.
You can erase an inactive host software version from the patient monitor. Erasing can prevent
the activation of a wrong software version by mistake.
1.
Log in to Webmin.
2.
Select
Configuration
>
Software Management
.
The
Software Management
window opens.
3.
Select the software that you want to erase and select
Next
.
The selected software activation window appears.
4.
Select
Erase
to erase the inactive software.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......