Concentration analysis 8
•
Examine the fitted curves in relation to the experimental data. Discard
samples where the fit is poor at one or both of the flow rates. As a general
guideline, the value reported for chi-squared should be less than about 5%
of the maximum response reached at the lowest flow rate (disregarding
the difference in units between chi-squared and response).
•
Examine the statistical significance of the reported concentration as
indicated by the standard error or T-value (see Section 9.4.2). The standard
error should be 10% or less of the reported concentration (T-value greater
than 10). Discard samples where the standard error represents more than
20% of the reported concentration (T-value less than 5), and treat results
where the standard error is in the range 15–20% of the concentration (T-
value 5–10) with caution.
•
Examine the values in the
Measured Conc
column. The operating range of
the method is approximately 0.05–5 µg/ml: values outside this range are
likely to be unreliable.
8.3.5
Fitting model
Evaluation of calibration-free concentration relies on fitting the data to a model
for 1:1 interaction (see Section 9.8.1) where the mass transport coefficient is
provided (through calculation from the diffusion coefficient, and flow cell
characteristics) and the analyte concentration is evaluated as a global variable.
Although this model is equivalent to the model for evaluation of 1:1 kinetics in
terms of the interaction description, the model for calibration-free
concentration analysis cannot be accessed through the model editor.
Notes:
The diffusion coefficient for the analyte at the analysis temperature is
calculated automatically from the value at 20°C. The value at 20°C is
provided as an evaluation variable when the assay is run (Section B.2).
Parameters for calculating the mass transport coefficient from the
diffusion coefficient are fixed in the software.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......