B40/B20 Patient Monitor
15-10
Smart InvBP and flushing
Flushing is performed to keep the lines open. It prevents blood from clotting and occluding the
lines and measurements. Infusion that is used for flushing goes through the dome into the
patient’s artery.
Two types of flushing are used simultaneously: flushing with continuous infusion and manual
flushing. Flushing with continuous infusion uses a higher pressure than the patient’s blood
pressure and contains Heparin to prevent blood from clotting. Manual flushing is always used
after having taken a blood sample, and every now and then to ensure that the lines remain
open.
The monitor detects the pressure used for infusion to flush the invasive blood pressure line.
There is a 40 second time-out for performing the flushing. The digits change to dashes, ---/--- ,
and there are no alarms. After 40 seconds, or when flushing is completed and patient’s pulse
detected, the pressure digits are displayed and alarms become active. If pressure values are
outside the alarm limits, an alarm is given.
The monitor functions the same way when blood samples are taken.
Checklist
Check that:
•
Invasive blood pressure transducer cable is plugged to the adapter cable, and this is
connected to the red connector in the module.
•
Pressure transducer is connected to the cable.
•
Patient catheter is connected to the pressure line.
•
There is no air in the transducer dome or catheter line and transducer is at mid heart
level.
•
Pressure transducer is zeroed.
•
Invasive blood pressure is selected for screen through
Monitor Setup
-
Screen Setup
.
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......