Invasive blood pressure
15-7
the heart rate calculation in a specified order: ECG (the lead with highest R-wave), pressure
(Art), and plethysmographic pulse waveform.
IBPX Alarm
Alarms can be adjusted in this menu or in the
Alarms Setup
menu which opens by pressing the
Alarms Setup
key.
Determining pressure values visually
By moving the horizontal cursor across the pressure waveform, you can get accurate pressure
values at selected points. This may be useful, for example, if the patient's breathing pattern is
irregular. The cursor is not available for pressures shown with a combined scale.
1.
Press the
IBP
key.
2.
Select
Px Setup - Px Cursor
.
3.
Move the cursor up or down by turning the Trim Knob. Every time the cursor is moved, the
time (hours and minutes) and pressure values are displayed on the screen. This way, you
can keep track of the changes made.
4.
You can remove the cursor by selecting
Remove cursor
. Note that if the cursor is not
removed, it remains visible in the Normal Screen.
Labeling channels
The label of the pressure channel sets its display scale, color, filter, alarm source and alarm
limits. The label descriptions are preconfigured.
To change the label:
1.
Press the
IBP
key.
2.
Select
IBP1 Setup
.
3.
Select
Label
.
IBP
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......