Non-invasive blood pressure
14-7
Starting
Note that the measurement unit may be mmHg or kPa. The unit is selected during
configuration through
Monitor Setup
-
Install/Service
-
Installation
-
Units
.
You can start the NIBP measurement using either the direct function keys, or from the NIBP
menu by pressing the
NIBP
key.
In the beginning of the measurement, sys and dia labels are replaced by the inflation limit
indication (
Adult/Pediatric, Neonatal
) for five seconds. The cuff pressure is displayed in the
mean pressure value field.
If motion artifacts are detected, the monitor automatically holds deflation until the motion
stops (maximum of 30 seconds). If the artifacts prevent proper measurement, a new
measurement starts automatically.
When the measurement is ready, you can hear a short beep and see the result numbers
flashing.
During measurement
•
Observe the cuffed limb frequently. Measurement may impair blood circulation. Intervals
below 10 minutes and STAT measurements are not recommended for extended periods
of time.
•
Make sure that the tubes are not bent, pressed or stretched. Measurement may be
impaired.
•
Blood pressure values may be affected by a change in the patient's position.
•
The presence of some arrhythmias during NIBP measurement may increase the time
required for the measurement.
NIBP
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......