Setting time zone
Time Zone
menu is enable only when monitor is not connected to network and discharge
paitent/haven’t admitted patient.
1.
Press
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Time Zone
•
Daylight Saving
: to turn on, turn off or set auto adjust daylight saving time (DST).
•
DST Offset Hour
: to set the hours offset for daylight saving.
•
DST Offset Minute
: to set the minutes offset for daylight saving.
•
DST Begins
: to set daylight saving begin time.
•
DST Ends
: to set daylight saving end time.
•
Time Sync
: to set time sync options,
Unity
is time sync to Unity Network,
NTP
is time
sync to NTP Server.
•
NTP Config
: to set NTP Server configuration.
-
GMT Offset Hours
: to set offset hours from GMT (Greenwich Mean Time).
-
GMT Offset Minutes
: to set offset minutes from GMT.
-
Primary Server IP
: to set primary NTP Server IP address.
-
Secondary Serv IP
: to set secondary NTP Server IP address.
NOTES:
−
DST Offset Hour
and
DST Offset Minute
menu are enable only when the
Daylight Saving
is set to
ON
or
AUTO
.
−
DST Begins
and
DST Ends
menu are enable only when the
Daylight Saving
is set to
AUTO
.
−
Time Sync
menu is shown only when monitor configuration is Unity+HL7.
−
NTP Config
menu is shown only when monitor configuration is Unity+HL7 or HL7.
NTP
Config
menu is enable only when
Time Sync
is set to
NTP
.
Monitor
Setup
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......