B40/B20 Patient Monitor
6-4
Changing alarm options
1.
Press the
Monitor Setup
key, select
Install/Service
and enter the password.
2.
Select
Installation - Alarm Options
.
•
Show limits
:
YES
(default) is show alarm limits in digit fields.
•
Show Audio ON/OFF
:
YES
is enable alarm silence options in the
Audio ON/OFF
menu in
Alarms Setup
.
NO
is the default.
•
Latching Alarms
:
YES
to keep alarm messages on screen until
Audio pause
key or
Alarm Reset
key pressed.
NO
is the default.
•
Breakthrough Alarm
: Select
Off
to close breakthrough alarms feature, select
Pedi
to
turn on breakthrough feature for new fatal alarms, select
Adult
to turn on
breakthrough feature for new fatal alarms, excluding Brady.
•
Min Alarm Volume
: Adjust minimum volume of audible alarm tone.
•
Reminder Volume
: Adjust the volume of the audible alarm reminder tone.
•
Alarm Tones
: Choose the
IEC
,
ISO
,
ISO2
(default) or
General
for alarm tone pattern.
•
Remote Control
: Select
active
to enable alarm remote control by central station.
Inactive
is the default.
•
NurseCall:
Select
NO
, high electrical level is exported from Nurse call connector
when there is medium or high priority alarm. Select
NC
low electrical level is
exported from Nurse call connector when there is medium or high priority alarm
.
NOTE: The
Show Audio ON/OFF
setting should be changed only by the system
administrator.
Other alarm settings (alarm limits, and alarm volume) can be changed in the
Alarms Setup
menu. To make the changes permanent, save them in user modes, see section
Monitor
Setup
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......