Technical Sheet

 

ref C1519B

XDI-XDIwin – 15/30J

SEMICONDUCTOR SENSOR 

Set-up Procedure: 292 D7C Issue Bv3

• 

NEW SENSORS ARE SUPPLIED READY TO CONNECT AND GO  
BUT NEED 15 MINUTES POWER STABILISATION. The following procedure is 
for full set up using a PC but onsite calibration without PC/LCD requires only 
sections 18 & 19.

Note:

 If a new cell is fitted then full calibration is required.

PC setup

This sensor will linearise 2 points, zero and mid-scale with an optional full scale point when using a PC.

Gas values below the mid-scale point are not as accurate because the sensor is none linear.

The Zero adjust is continuous during power up inhibit time and PCB fault LED flashes to indicate this, do 
not apply any gas at this time because Zero adjust will try to offset the measurement. Thereafter Zero 
adjusts 1 second every minute. D1 flickers orange at the zero point.

Voltage measurements are made wrt AG unless otherwise specified. 

1.  Insert jumpers J20 and J6 position SO for 4-20mA source output
2.  Turn sensor voltage potentiometer RV4 fully anticlockwise (minimum voltage)
3.  Connect sensor to J2 terminal (W) white (P) pink (Y) yellow
4.  Connect 24v + and 0V to J10, short the 4~20mA terminal by inserting test link at J9
5.  Set the sensor voltage to 5.0V (4.0 for NH

3

)measured at SV and 0V, adjust by turning RV4 

clockwise 
DO NOT EXCEED THE REQUIRED VOLTAGE OTHERWISE PERMANENT DAMAGE WILL OCCUR

6.  Insert jumper G2 for a nominal gain value (G1 is lowest gain, G5 is highest or a combination G1 

to G5)

7.  Connect a PC to J3 with and RS232 Combi adaptor (part no. 160510 and lead part no. 160515) 

and Hyperterminal 4800/1stop/8bit  and initialise the sensor using (C) calibration mode, then shift 
+ ($) command from the keyboard. The default gas is Ammonia 1000 ppm

8.  The typical Hyperterminal display is (also see appendix 1)
 

27 Ammonia 1000.0ppm

 

O H L D F I

 

^ ^ ^

 

- - - - - -    PPM=  0 A=80 R=80 Adj=497 PInh=52 adj=4 CG=314.0 CGADC=385

 PPM

  

is the gas value

 

A

 

is the processor raw measurement (0 to 1000)

 

R

  

is the zero point of the sensor (100 +/-2)

 Adj

 

is the adjustment value to autozero 0 to 1000 but 500 is ideal.

 

Pinh

 

is the number of seconds left of powerup inhibit (autozero is active during this time)

 adj 

 

is the number of seconds until a 1 second autozero will occur.

 

CG

 

is the calibration gas value PPM.

 

CGADC 

  is the number A must reach to display the value CG as the correct PPM gas reading

Summary of Contents for XDI-XDIwin 15/30J

Page 1: ...e output 2 Turn sensor voltage potentiometer RV4 fully anticlockwise minimum voltage 3 Connect sensor to J2 terminal W white P pink Y yellow 4 Connect 24v and 0V to J10 short the 4 20mA terminal by inserting test link at J9 5 Set the sensor voltage to 5 0V 4 0 for NH3 measured at SV and 0V adjust by turning RV4 clockwise DO NOT EXCEED THE REQUIRED VOLTAGE OTHERWISE PERMANENT DAMAGE WILL OCCUR 6 In...

Page 2: ...ied auto zero is in operation and LED D should be flashing orange Check the voltage between Ag and S16 to be 450mv approx With J9 test link inserted the 4 20mA output signal measured at TP8 to TP9 should be 4mV 19 Span ensure that auto zero is not within 30 seconds indicated by the MPV LED short flash then apply test gas typically 50 for 2 minutes and use the UP and DOWN buttons to adjust the 4 20...

Page 3: ... HIGH ALARM is next followed by OVER RANGE alarm The direction of the alarms is displayed as for rising and v for falling but these can be changed using left and right magnets together Appendix 1 Hyperterminal Display OHLDFI O H and L represent the Over Range High and Low alarms respectively v shows the direction of alarms D indicates if a duplicate address is detected F indicates a fault I indica...

Page 4: ...dified at any time without prior notice Technical Sheet ref C1519B XDI XDIwin 15 30J SEMICONDUCTOR SENSOR AMMONIA Set up Procedure 292 D7C Issue Bv3 Figure 1 Figure 2 Figure 3 Appendix 2 The semiconductor sensor can react to gas flow and humidity When gas is removed from the sensor the displayed PPM gas value may rise and take some time to return to normal Autozero is therefore delayed by 2 minute...

Reviews: