6 Maintenance
Functionality, reliability and safety characteristics of ultrasound therapy devices are only guaranteed if
properly used in accordance with the operating instructions. Safety control, maintenance, repair and
modifications shall be carried out by the manufacturer or the service agents authorized by him. In case of a
failure, parts which influence the safety of the device shall be replaced only by original spare parts of the
manufacturer. The electric installation shall be carried out in accordance with the requirements of VDE/IEC.
The device does not contain any parts which need maintenance work done by the user.
6.1 Safety controls
The device is subject to the provisions of the MDD. The safety controls have to be carried out on the basis
of this directive. Thereby, the operator regulation has to be especially observed.
Irrespective of the legal rules or beyond the scope of the MDD, it is recommended to have the device
checked by the manufacturer or by a service agency authorized by him at 12-months intervals.
The check shall consist of at least the following:
•
Electrical safety check in accordance with the test plan of the manufacturer,
•
Check of the device in respect of external integrity,
•
Check of all display and operating elements in respect of damage,
•
Check of all inscriptions in respect of legibility.
6.2 Cleaning, disinfection and care
For cleaning and disinfection of the device and its accessories there should not be used any agents
containing higher portions of phenol derivates, alcohol, compounds of chlorine or peracetic acid. It is
recommended to use disinfectants on aldehyde basis.
The device is not suited for heat sterilization or for sterilization with gases.
Warning!
The device and the ultrasound probes are suited for wiping disinfection. It has to be observed that no liquids
enter the device or the ultrasound probes. Never shall the plug or socket get wet. For cleaning or disinfection
the device must not be dizzled.
!!
Note!
After each treatment session, be sure to thoroughly clean the resonance plate of the treatment probe to
prevent oxidation.