14
SONOSTAT
135
gbo Medizintechnik AG 2004
Version 1.3
Power for
5
cm
2
ultrasound probe
2.5 cm
2
ultrasound probe
1 MHz
0 to 15 W
0 to 3 W/cm
2
0 to 15 W
0 to 3 W/cm
2
3 MHz
0 to 7.5 W
0 to 1.5 W/cm
2
0 to 7.5 W
0 to 1.5 W/cm
2
Table 5:
Range of ultrasound power
Figure 9
:
Ultrasound probe small key
Figure 10
:
Ultrasound
probe large key
With the
ultrasound probe small key
or
ultrasound probe large key
one of both connected probes can be
selected. The appropriate key is only active (LED is lit) if an appropriate ultrasound probe is connected. If a
probe is selected, the LED of the key flashes. The keys are locked during the treatment.
!!
Note!
•
The coupling status of the ultrasound probe is indicated by the LED in the ultrasound probe
housing. The LED flashes in case of insufficient coupling.
•
If the ultrasound probe is not contacted to the patient during therapy, no therapy time will pass.
After reestablishing the contact to the patient the therapy time will pass further. The ultrasound
probe can be removed at any time. Exception: You changed the linkage in the menu Settings,
see
3.2.6 Therapy time
The therapy time can be set in minute steps between 1 and 15:
1.
Press
the
therapy time key
.
2.
Set the desired therapy time
by means of the
modification keys
.
Figure 11:
Therapy time key
!!
Note!
•
The therapy time can be modified during the treatment.
•
By pressing the
therapy time key
the LED starts flashing (active state). After a timeout of 10 s the
active state will be left.
•
At the end of the therapy time the ultrasound power is automatically switched off:
∗
On switching off the ultrasound power
the
intensity
has automatically reverts to
„0.0”
and the
triad gong is sounded 3 x.
∗
Parameter modifications of the last therapy are retained.
∗
The therapy can be continued by setting of a new therapy time and turning on the intensity.