32
HiToP 1touch
gbo Medizintechnik
AG
Version 1.2
9
Technical Data
Mains voltage and
frequency:
100 – 240 V, 50-60 Hz
Power consumption:
max. 45 VA
Output current:
max. 310 mA
Output voltage:
max. 76 V
Output power:
max. 5000 mW
Patient resistance:
30 – 1500 Ω
Applied part
Contact surfaces at the electrodes.
Mode of operation:
continuous operation
MDD device class:
IIa
Protection degree:
I acc. to IEC 60601-1
Protection class:
BF acc. to IEC 60601-1
Protection against
ingress of: liquids
IP X0
Dimensions:
max. 27 cm x 24,5 cm x 6 cm (W x H x D)
Weight:
2,44 kg without accessories
Color:
aluminium natural anodized and grey RAL 7016
Display:
10,1“ TFT LCD touch screen
Environmental
conditions:
operation of the
device:
temperature range +10 °C ... +35 °C
relative humidity of air 30 ... 75 %
transport and
storage:
temperature range +5 °C ... +50 °C
relative humidity < 90 %, non
condensing
Current types:
carrier frequency 4096 –32768 Hz Sinus wave
low frequency
0.1 Hz – 200 Hz
accuracy
±
10 %
By request of technical personnel gbo Medizintechnik AG can offer spare part lists and circuit diagrams.
The appliance inlet is used for all pin disconnection from the mains.
gbo Medizintechnik AG reserves the right to modify design and specifications without
prior notice.