System default:
DAR=
1-min insulation resistance
30-sec insulation resistance
5) Automatic Measurement of Dielectric Absorption Ratio (DAR) and Polarization Index (PI)
When the measurement time reaches the set time of dielectric absorption ratio (DAR), the tester will
automatically calculate and display the DAR value, as shown in Picture 21.
Other time can be set in the "Settings" option to calculate the DAR value.
Picture 21
When the measurement time reaches the set time of polarization index (PI), the tester will automatically
calculate and display the PI value.
System default:
PI=
10-min insulation resistance
1-min insulation resistance
Picture 22
Other time can be set in the "Settings" option to calculate the PI value, as shown in Picture 22.