background image

No.131TP

Instructions for

Vinyl Chloride Detector Tube

FOR SAFE OPERATION :

  Carefully read this manual and the instruction manual of your Gastec Gas Sampling Pump.

CAUTION : If you do not observe the following precautions, you may 

suffer injuries or damage the product.

1. When breaking the tube ends, keep away from eyes.
2. Do not touch the broken glass tubes, broken pieces and reagent with bare hand(s).

NOTES :  For maintaining performance and reliability of the test 

results, observe the following.

1. Recommend to use Gastec Gas Sampling device Model GSP-300FT-2 (if not available use 

the air sampler of equivalent to sample for 100 mL/min) together with Gastec Detector Tubes 
only for the purposes specified in the instruction manual of the detector tube.

2. Use this tube within the temperature range of 0 - 40

 (32 - 104

).

3. Use this tube within the relative humidity range of 0 - 90%.
4. This  tube  may  be  interfered  with  by  the  coexisting  gases.  Please  refer  to  the  table 

"INTERFERENCES" below.

5. The shelf life and storage condition of the tube are marked on the label of the tube box.

APPLICATION OF THE TUBE :

  Use this tube for detecting Vinyl Chloride in the air or in industrial areas and for determining 

the environmental atmospheric condition.

SPECIFICATION :

  (Because of Gastec's commitment to continued improvement, specifications are subject to 

change without notice.)

Measuring Range

0.2 – 3.0 ppm

3.0 – 9.6 ppm

Sampling Rate

100 mL/min

50 mL/min

Correction Factor

1

3.2

Sampling Time

10 minutes

Detecting Limit

0.08 ppm

Colour Change

Yellow  

  Reddish purple

Reaction Principle

CH

2

 : CHCl 

 I

2

O

5

 H

2

SO

4

  

  HCl, Cl

2

HCl, CI

2

 

 Base  

  Chloride

Coefficient of Variation : 10% (for 0.2 to 1.0 ppm), 5% (for 1.0 to 3.0 ppm)

**Shelf Life : Please refer to the validity date printed on the tube box.

**Store the tubes in a dark and cool place.

CORRECTION FOR TEMPERATURE, HUMIDITY AND PRESSURE :

Temperature : 

Correct for temperature with the table below.

Temperature 

(

)

(32)

5

(41)

10 

(50)

15

(59)

20

(68)

25

(77)

30 

(86)

35

(95)

40

 (104)

Correction Factor

3.00

2.15

1.60

1.20

1.00

0.90

0.80

0.70

0.60

Humidity : 

No correction is required for 0 - 90%R.H.

Pressure : 

To correct for pressure, use the formula below :

   

   Tube Reading (ppm) 

×

 1013 (hPa)

   

 Atmospheric Pressure (hPa)

MEASUREMENT PROCEDURE :

If automatic air sampler Model GSP-300FT-2 is used

1. Prior to operation please confirm if black colour inlet rubber tube holder is equipped with the 

sampler.

2. Break both end of the tips of the detector tube by the tube tip holder supplied. Connect the 

detector tube to the pump with arrow ( 

 ) on the tube pointing toward pump.

3. Set the flow meter at 100 mL/min and timer to “10 minutes” of the sampler.  Press the start 

switch of the sampler to start the sampling.

4. After the sampling, remove the detector tube from the sampler.
5. Read the concentration level at the interface where the stained reagent meets the unstained 

reagent.

6. For measurements higher than 3.0 ppm, prepare a fresh tube and set the flow meter at 50 

mL/min and timer to “10 minutes” of the sampler.  Press the start switch of the sampler to 
start the sampling.

7. If necessary, multiply the readings by the correction factors of temperature and atmospheric 

pressure.

INTERFERENCES :

Substance   

Concentration

Interference

Interference gas only

Hydrogen Chloride

0.2 ppm

+

Reddish purple at 0.2 ppm

Chlorine

0.04 ppm

+

Reddish purple

Ethylene

55 ppm

-

No discolouration

1,2-Dichloroethane

10 ppm

No

No discolouration

Tetrachloroethylene

50 ppm

No

Reddish purple at 500 ppm

Trichloroethylene

0.04 ppm

+

Reddish purple

Toluene

1.0 ppm

-

No discolouration

Benzen

0.8 ppm

-

No discolouration

  This table of interference gases primarily expresses the interference of each coexisting gas in 

the concentration range, that is equivalent to the gas concentration. Therefore, the test result 
may show positive results due to other substances not listed in the table. If more information 
is needed, please contact us or our distributors in your territory.

ppm

131TP

0.2

0.5

1.0

1.5

2.0

2.5

3.0

Detecting  Layer

Oxidising  Agent

Reviews: