Gast 0211 Series Operation & Maintenance Manual Download Page 1

KEEP THIS DOCUMENT FOR FUTURE REFERENCE

    

 70-180  G325PL  (Rev. C)

0211 SERIES LUBRICATED LABORATORY

VACUUM PUMPS & COMPRESSORS

OPERATION & MAINTENANCE MANUAL

CONTENTS:

General Information and Installation ....................................................................................................... 2

Operation and Maintenance ..................................................................................................................... 3

Exploded View and Parts Ordering Information ..................................................................................... 4

Shutdown Procedures, Service Kit Installation, and Troubleshooting Guide ......................................... 5

Warranty and Authorized Service Facilities ............................................................................................ 6

Visit us at our website

www .gastmfg.com

Summary of Contents for 0211 Series

Page 1: ...SORS OPERATION MAINTENANCE MANUAL CONTENTS General Information and Installation 2 Operation and Maintenance 3 Exploded View and Parts Ordering Information 4 Shutdown Procedures Service Kit Installation and Troubleshooting Guide 5 Warranty and Authorized Service Facilities 6 Visitusatourwebsite www gastmfg com ...

Page 2: ...t flow restriction use pipe and fittings that are same size as or larger than threaded ports Note that ports are not designed to support plumbing m WARNING Improper wiring can result in electric shock All power to motor must be de energized and disconnected before servicing Wiring Refer to motor nameplate for wiring diagram Building must conform to all required safety codes Check that voltage at p...

Page 3: ...o a stuck vane caused by dirt or excessive lubrication mDANGER Do not use kerosene or other combustible solvents to flush unit Hazardous vapors from flushing fire or explosion could result and cause personal injury or death m WARNING Flush unit in a well ventilated area Eye protection is recommended Keep face away from exhaust port Follow these steps to flush this unit 1 Disconnect from power sour...

Page 4: ...s a l G 1 3 8 5 B A 1 1 y l b m e s s A r e t l i F l i O 1 J 0 8 6 A A 2 1 t e k s a G r e v o C 1 1 9 6 A A 3 1 t l e F r e t l i F 3 B 3 8 6 A A 4 1 t l e F r o t a r a p e S 1 8 9 6 A A 5 1 k c i W r e l i O 1 9 9 6 A A 6 1 r a J s s a l G 1 3 4 0 9 6 A A 7 1 e n a V 4 B 5 4 3 A A 8 1 y l b m e s s A t r o p p u S t o o F 1 6 3 1 C A 9 1 y l b m e s s A g u l P h c t i w S d r o C 1 6 9 8 A A ...

Page 5: ...unit under vacuum 6 Disconnect unit from power source 7 Plug open ports to prevent dirt or other contaminants from entering unit Unit is ready for storage SERVICE KIT INSTALLATION Note Gast will not guarantee performance of a field rebuilt unit Return unit to a Gast Authorized Service Facility or perform following rebuild procedures Service kit contains four vanes filter element element gasket jar...

Page 6: ...RECT OR CONSEQUENTIAL DAMAGES OF ANY KIND however arising including but not limited to those for use of any products loss of time inconvenience lost profit labor charges or other incidental or consequential damages with respect to persons business or property whether as a result of breach of warranty negligence or otherwise Notwithstanding any other provision of this warranty BUYER S REMEDY AGAINS...

Reviews: