7 Installation GC-IMS Device
7.1 Installation Requirements
The following requirements must be fulfilled by the customer:
Location of Installation
•
Available space of 600 x 700 x 300 mm (W x D x H)
•
Ambient temperature of 5
– 40 °C
•
Humidity: 0-90 % RH, non-condensating
Electricity
•
Electricity supply free of interferences
•
Power Supply of 230 V ± 10%, 50- 60 Hz ± 1%
Gas supply
•
Nitrogen (Quality 5.0 (99,999%) or Synthetic Air (Quality 5.0 (99,999%)
•
Stainless steel pressure reducer with adjustable pressure range of 3
– 6 bar
and 3 mm Swagelok connector
Safety
•
Availability of exhaust system for device exhaust gas tubes (Gas out and
Sample Gas out)
Computer
•
Computer with current Microsoft Windows operating system
•
Administrator right to install G.A.S. software