GC-IMS
– User Manual
79/214
Sound:
The frequency of the warning sound can
be set.
Compression:
Compresses the measurement data to zip
6.7.3.1 Snapshot Window
The instrument performance and status can be monitored using recorded snapshots.
It is an automatic comparison of current instrument settings versus the operator
defined target settings which helps the operator to assess the readiness of the
system for starting a measurement. This automatic monitoring can recognize among
other a System contamination, insufficient gas quality or system leaking. For this
purpose, the following parameters are constantly cross-checked against the target-
settings:
•
The carriergas pressure - EPC2 pressure (kPa)
•
The height of the Reaction-Ion-Peak (RIP) - Rip-height (V)
•
The position of the Reaction-Ion-Peak (RIP) normalized to normal pressure
–
Rip Pos at 101.33 kPa (ms)
•
The full width at half maximum of the Reaction-Ion-Peak - FWHM (ms)
•
The temperature values T1-T6
•
The gasflow and pressure of Driftgas (EPC1) and Carriergas (EPC2)
The user can change the default factory settings for his needs. Error messages are
displayed in the Error Information Window.