![GAS GC-IMS User Manual Download Page 69](http://html1.mh-extra.com/html/gas/gc-ims/gc-ims_user-manual_3647276069.webp)
GC-IMS
– User Manual
69/214
P1:
The sample flow can be set in percent of the pump power
(0- 100
%)
Selected program action line
End of Program line
INFORMATION!
The EPC2 maximum flow (carriergas) of 150 ml/min is restricted by
the installed column dimemsions. It only can be attained if the
standard column is installed (15 m length, 0.53 mm inner
diameter).
Using Columns with other dimensions can linited the
EPC2 maximum flow.
A value here determines the new state of the respective device component. A vertical
line “|” indicates that the respective device component is not involved in this action.
The last row in this action list contains horizontal lines only. It indicates the end-of-
program. The point of time of this end-of-program marker can be increased. This
determines the duration of the program run.
Figure 45: Selected Program Window Control Panel
With the buttons in the Selected Program Window Control Panel the currently
displayed program can be created, modified or deleted. A newly created action is
positioned after the currently selected action. Pressing the
Copy
button copies the
selected action and orders the copy behind the selected action. To maintain the
chronological order of the actions the system may rearrange the actions by ordering
them according to their numbers later on. By pressing the
Del
button the currently
selected action is deleted. Pressing the
Edit
button or the
New
button opens the Edit
Action dialogue.
Summary of Contents for GC-IMS
Page 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Page 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Page 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Page 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Page 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...