![GAS GC-IMS User Manual Download Page 48](http://html1.mh-extra.com/html/gas/gc-ims/gc-ims_user-manual_3647276048.webp)
GC-IMS
– User Manual
48/214
Horizontal Control
buttons
Moves the horizontal position of the
display area on the screen left or right.
Zoom Control buttons
Reduces or enlarges the view of the
display area on the screen
6.1.5 Low/High Pressure Control
The device pressure is monitored. Two types of error are defined:
6.1.5.1 Low Pressure Error
Figure 20: Operating Interface - Low Pressure Alarm Box
1. After 10 seconds a dialogue box is displayed and an acoustic alarm sounds.
2. After 5 minutes all temperature controllers are switched off.
3. When the pressure is reached again the alarm is switched off and all
temperature controllers switch themselves on automatically