23
5.3
Workflow Cleaning
1
Typical View of the Start
Page after successful set-up
and
power-up
of
the
instrument. The instrument
would be ready, the key
instrument
default
parameters
have
almost
reached their setpoints (see
red arrows)
Contamination is seen
(red circle), the inst-
rument needs cleaning.
Ideally do a Cleaning over
night
2
Open:
System
>
Modes
>
Cleaning
Mode
Chose between
High Flows
or
Reduced Flows
It is recommended to use
high flows because the
Cleaning is more efficient.
Chose Reduced Flows if
Nitrogen saving is necessary.
Select the
Cleaning duration
Click Start to start the
Cleaning cycle