Operation
36
Operation Manual LVS 105 T
– 10 ef
+
6.2.5
Operating mode 5 Self-cleaning
The
self-cleaning operating mode
handles the purging and drying of the pump and the
pneumatic supply lines.
PARAMETER
VALUE
Unit
Cleaning Time
120
s
Parameter
Values from/to
Resolution
Default
Unit
Cleaning Time
1…600
1
120
s
The ventilation valve is opened and, due to the flow rate generated by the vacuum pump, the
pump and the pneumatic supply lines are purged in the specified time. The integrated vac-
uum pump operates at the maximum rotational speed.
ATTENTION
►
The application is ventilated by the self-cleaning procedure.
►
If no other gas type is connected to the inert gas connector of the ven-
tilation valve, cleaning is done with air.