Operation
32
Operation Manual LVS 105 T
– 10 ef
+
6
Operation
6.1 Bringing into service
ATTENTION
►
Before starting up the device, read the safety instruction; see section2
Safety.
1. Connect the device as specified and shown in the figures; see sections
2. Switch on the device at the main power switch I/O on the back. After a brief initialization
routine, during which STARTING is shown, the device is ready for operation.
6.2 Operating modes and system settings
6.2.1
Operating mode 1 Manual
In the manual operating mode, the setpoint of the pressure is readjusted by the system.
PARAMETER
VALUE
Unit
Set Pressure
1000
mbar
Auto Stop Time
00:00:00
hh:mm:ss
Vent at Finish
Yes / No
Parameter
Values from/to
Resolu-
tion
Default
Unit
Set Pressure
0…1100
1
1000
mbar
Auto Stop Time
00:00:00…23:59:59
01:01:01
00:00:00
hh:mm:ss
Vent at Finish
Yes / No
No
The pressure setpoint can be specified as a parameter or adjusted in real time while the op-
erating mode is active using the operating elements; see section
display. Both the regulated pressure drop as well as the pressure increase are possible.
The pressure increase requires, depending on the application, a ventilation of the system.
The latter must be released in the vacuum controller's system settings (see section