RC1xx LED Lighting Controllers - User Manual
—
31
—
Numeric parameters are separated by a comma (
,
). Time periods are
specified in milliseconds by default, but you can add ‘s’, ‘ms’, or ‘us’ to the
parameter to specify seconds, milliseconds, or microseconds. Current is
specified in amps by default, but you can add ‘A’ or ‘mA’ to the parameter
to specify amps or milliamps.
Note:
All parameters are specified in the ‘USA/UK’ format, where a half is
written as ‘0.5’ and not ‘0,5’. Examples of acceptable parameters are shown
below:
0.1
0.1 milliseconds
200us
200 microseconds
0.1s
0.1 seconds
100mA
100 milliamps
2.45A
2.45 Amps
2.3
2300 milliamps, 2.3 Amps
10.3.3 General commands
Any changes you make using Ethernet commands are not saved
permanently until you send the
AW
command.
Report the RC120 firmware version
This command returns the firmware version running on the RC120.
For example, typing:
VR
returns
RC120 (HW101) V001
Save the settings to memory
AW
This command saves the settings to non-volatile memory. If you
power cycle the RC120 before sending the
AW
command, all changes
are lost and the unit returns to the values set before the last
AW
command.
Set the option flags
This command sets the flags for error detection, trigger sense, and
SafeSense:
RE1,m