CL190/CL191 Lens Controller - User Manual
The following examples illustrate how these commands are used:
RN1,1,100
Set as continuous square wave at 10Hz.
RN1,8,50
Set a continuous triangle wave at 20Hz.
RN1,11,10s
Set a triggered sine wave which runs once
for 10 seconds.
RU1,2,0
RN1,10,200
Set a continuous sawtooth wave at 5Hz
from 0 to 2 diopters, with a slow fall and fast
rise.
RU1,2,0
RN1,10,200
Set a continuous sawtooth wave at 5Hz
from 0 to 2 diopters, with a slow rise and fast
fall.
9.7
Programmable step commands
Use this command to set up a sequence of steps where different optical
powers can be set up for a programmed period of time.
Set the number of steps in a sequence
RX1,0,n
Where:
n
= the number of steps (0 to 20)
Set a step point
RX1,1,e,f,t
Where:
e
step number (0 to [number of steps] - 1)
f
focal power in diopters
t
time at this step in milliseconds
Add a step to the sequence
RX1,2,f,t
Where:
f
focal power in diopters
t
time at this step in milliseconds
Add a step to the sequence using the current focus
RX1,3,t
Where:
t
= time at this step in milliseconds
—
29
—