Working procedure AA MTK 70_03
Ambit:
Ganshorn Medizin
Electronic GmbH
Technical Support
Working procedure for PowerCube Ergo /SSO Rev E
Revision:
Rev01 30.11.2012
Page 3 von 7
8. Electronics
Functional check of the switching and measurement channels in COM98IO:
Check the pressure sensors with pos. und neg. pressure; the higher differences have
to be noted.
Test procedures for gas sensors have to be carried out with running suction pumps
and complete tubing.
Attention: While adjustment do not breath into the system!!!
channel
Ergo
0
(SpO
2
)
1
---
2
---
3
Pressure(amb)
4
---
5
Temp.
6
---
7
---
8
PF1
9
---
10
CO
2
11
(Puls Digi.)
12
O
2
13
(Puls Ana.)
14
---
15
---
switch
Ergo
0
Pump Ergo
1
Cal. Valve
2
---
3
---
4
---
5
---
6
---
7
---
Switching channels
Check all used switching channels.
Flow PF1 Offset
Check Flow PF1 in COM98 I/O.
Adjustment:
Trim Flow PF1channel with potentiometerR8 on Analog-Modul 1.3
Flow PF1 Gain
Impinge pressure to sensor K7 and check value in COM98I/O.
Adjustment:
Impinge pressure and adjust Channel Flow PF1with potentiometer R23
on Analog-Modul 1.3