Working procedure AA MTK 70_03
Ambit:
Ganshorn Medizin
Electronic GmbH
Technical Support
Working procedure for PowerCube Ergo /SSO Rev E
Revision:
Rev01 30.11.2012
Page 2 von 7
Application area:
PowerCube-Ergo with Art.No: 033370210, 033370238
For options further checks might become necessary
Basic remark:
This check procedure must be carried out within a preventive maintenance and as
checkout procedure after repair.
1. Examinee
Enter in the checklist the serial numbers of all parts that are checked.
2. Check conditions
The check conditions are the same as the measurement conditions.
Avoid ambient condition fluctuations during at least 24 h before the check.
Check, if actual ambient conditions correspond to ambient condition in the LF8-
Calibration-program/ambient condition
3. Test equipment
Please note the requirements for test equipment. Use only test equipment with the
allowed measuring range and which are calibrated accordingly to their manufacturer
’s
guidelines (see chapter
„Necessary tools and equipment“)
4. Calibration gas
The gas specifications and requirements for checks are the same as for calibrations
during running operation. The actual values have to be entered correctly in LF8.
5. Deviations, establishments and evaluation
Experienced deviations which can not be repaired immediately must be noticed.
Differences can cause that some parts of the system or the complete system is not
allowed to use till restoration since carried out. If necessary the complete check must
be repeated. The evaluation shows the current state of the tested system.
6. Visual check
Perform visual check of the complete system for damages, failures, contamination,
wrong / insufficient connections (tubes, connectors) etc.
7. Software settings
Before the check, assure correct software settings (ambient conditions, calibration gas
etc.)