Fixate the styli with the respective screws in the retainers
of the lower face bow. The cable of the stylus should
hang downwards after it is fixated.
If the system is mounted properly and the lower jaw is in
centric position, the stylus tip should be located in the
area of the upper back quarter of the flag.
The styli should be positioned in a way that their axis protrude by about 1 cm to the outside.
This range of movement is, among other things, required for mediotrusive movements.
The circuit between stylus and flag should always remain closed, even when the system is dis-
mantled after use and the individual parts are stored in their packages. Improperly, jerkily break-
ing the circuit could lead to faults or disturbances in the system, e.g., broken cables.
Avoid touching the measuring surface of the flags and the stylus pins during the measurement
procedure and preparatory work. This also applies to the patient.
3.6
Setting up the CADIAX system
Plug the connector of the foot switch into the designated
jack at the backside of the device.
Connect the device to a free USB-port of your PC.