CADIAX compact
48
3.5.4
Mounting the writing bow
The two axis locators are inserted into the holes for the
left and right stylus and clamped firmly.
The side arms of the lower face bow are pushed com-
pletely outwards, then slightly fixated. The patient holds
the lower jaw closed and in centric (retral) position. The
double clamp is pushed onto the clutch.
Make sure not to loosen the clutch and check once more
that it fits correctly.
The side arms are moved, so that the tips of the axis loc-
ators slide into the bores on the flag adapters of the ex-
tension set, which mark the anatomical axis point. This
fixes the positions of the styli on the anatomical axis,
and also guarantees a correct distance to the flag.
The patient or an assistant lightly presses both guides in-
wards.
The writing bow is now adjusted so that it is parallel to
the upper face bow. Both fixing clamps of the side arms
are now carefully clamped firmly.