![Gamma Vacuum DIGITEL Instruction Manual Download Page 17](http://html.mh-extra.com/html/gamma-vacuum/digitel/digitel_instruction-manual_3370463017.webp)
7.
Touchscreen user interface
7.1.
Home screen
Tap the screen to show "Turn ON" and "MENU."
"MENU" takes you to NEG recipe selection, custom recipe setup, and display
setup.
Figure 7. Home screen
Line one indicates TSP status. Following are possible statuses:
▪
Initializing
▪
OFF
▪
Armed, X/Y min
▪
Armed, Interlock - Active
▪
Armed, Waiting - In use
▪
ON X/Y A
The ‘OFF’ status indicates TSP programmed mode is not running.
Armed status indicates TSP programmed mode is running, but actual TSP
filament sublimation is not in progress at this time.
For example, ‘Armed, 2/15 min’ status indicates two minutes has passed
from last sublimation cycle where period between sublimation cycles is set
to 15 minutes.
The ‘Armed, Interlock -Active’ status indicates TSP programmed mode is
running, but sublimation is not possible because TSP interlock is not
satisfied.
The ‘TSP 1 Armed, Waiting -In use’ status indicates TSP 1 programmed mode
is running, but sublimation is waiting for TSP 2 sublimation to complete.
The ‘ON 40/45A’ status indicates filament sublimation is in progress. The
filament current is ramping up and currently at 40 Amps, where target
sublimation current level is set to 45 Amps.
Line two indicates current TSP setup. See
on page 18 for details.
Page 17
900035_B
900035_B - Touchscreen user interface