Step 2: If ‘Output state = Inverted’ multiply current log value in step 1 by (-1)
Step 3: Add offset value (-9 + 11 = 2 V)
The analog output pin will read 2 V.
Figure 8. Analog output screen
9.7.
Digital input setup
Total of four digital inputs on connector J104.
Digital input 1 (J104, pin 22)
Digital input 2 (J104, pin 23)
Digital input 3 (J104, pin 24)
Digital input 4 (J104, pin 25)
By default all digital inputs are turned off (function set to off).
Available functions:
▪
OFF
▪
HV interlock
▪
HV switch
▪
TSP 1 interlock
▪
TSP 2 interlock
HV interlock function
Ground pin to satisfy interlock. If interlock is not satisfied, HV cannot run.
HV switch function
Ground pin to turn on HV, otherwise HV is off.
TSP 1 interlock function
Ground pin to satisfy TSP 1 interlock. If interlock is satisfied, TSP 1 cannot
run.