General Alarms and Troubleshooting
12-52
COBE
®
Spectra™ Apheresis System •
Essentials Guide
RBCs LOST!
Alarm LED is red. Pumps stop.
During Run mode, to override the alarm for 4 minutes press
CONTINUE
. During Rinseback mode, to completely override the
alarm, press
CONTINUE
.
This alarm occurs if RBCs are detected at the start of Single-Needle Run
mode and the RBC detector later determines that the RBC level has
dropped below its initial level.
Restarting centrifuge.
Warning LED is Yellow.
RBCs LOST!
Check for open saline clamps.
CONTINUE
Probable Cause
Action
Access or return saline lines left open
Close the access and/or return saline line roller clamps. Press
CONTINUE
.
Line incorrectly positioned in the RBC line valve
Ensure the RBC line is placed correctly in the RBC/plasma line valve (pressed firmly in the bottom). Press
CONTINUE
.
RBC detector malfunction
Procedure can continue. The Spectra system uses time-out backup to close the waste divert valve and open
the return line valve.
Verify this action. If it does not take place, manually close the waste divert valve and open the return line
valve.
__.__
___.__ ___.__ ___.__ ___.__
____
Restarting centrifuge.
Probable Cause
Action
CONTINUE
pressed to restart the centrifuge
Press
CONTINUE
when the “Centrifuge up to speed.” warning appears.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...