General Alarms and Troubleshooting
12-42
COBE
®
Spectra™ Apheresis System •
Essentials Guide
NO SALINE SEEN AT INLET AIR SENSOR!
Alarm LED is red. Pumps stop.
Press
CONTINUE
to retry the priming inlet test.
This alarm occurs only during Prime mode if the inlet air sensor still
detects air after a specific time period.
NO SALINE SEEN AT INLET AIR SENSOR!
Check sensor for loading.
CONTINUE
Probable Cause
Action
Inlet air chamber not placed in the inlet air detector
Ensure the inlet air chamber is placed correctly in the inlet air detector. Press
CONTINUE
.
Saline container not connected
– or –
Access saline line clamp not open
Ensure the saline container is connected and that the access saline line roller clamp is open. Press
CONTINUE
.
Sensors not detecting fluid in inlet air chamber
Reposition the inlet air chamber. Ensure it has good contact with the sensor and the filter is below the sensor.
Press
CONTINUE
.
Centrifuge pressure sensor tubing out of place
Reinsert the centrifuge pressure sensor tubing. To reprime the disposable tubing set, press
CHANGE MODE
,
then press
2
to select “Prime.”
Return line valve unloaded
Reload the return line valve. Press
CONTINUE
.
Plasma line inserted in RBC valve
Ensure the RBC line is in the RBC valve. Press
CONTINUE
.
Pump or sensor malfunction
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm
message.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...