General Alarms and Troubleshooting
12-36
COBE
®
Spectra™ Apheresis System •
Essentials Guide
FLUID LEAK IN CENTRIFUGE!
Alarm LED is red. Pumps and centrifuge stop.
To override the alarm for 60 seconds during Prime or Rinseback mode,
press
CLEAR
.
If you are certain there is no leak, press
CONTINUE
. This restarts the
centrifuge and continues the procedure.
FLUID LEAK IN CENTRIFUGE! Check channel.
Dry fluid from sensor.
CONTINUE
Probable Cause
Action
Leak in channel
– or –
Fluid spilled on sliding cover and/or
centrifuge
To discontinue the procedure, perform the following actions:
1
After the centrifuge stops spinning, press
UNLOCK COVER
.
2
Slide back the centrifuge cover and lower the centrifuge door.
If manual Rinseback is an option:
1
Dry the surface of the fluid leak detector with a dry cloth. Discard the cloth in a biohazard disposal container.
2
Raise the door and close the cover.
3
Remove the alarm message and return to the Run screen by pressing
CLEAR
two or more times.
Note:
DO NOT press
CONTINUE
because this restarts the centrifuge. If you press
CONTINUE
, immediately
press
STOP SPIN
. If the “FLUID LEAK IN CENTRIFUGE!” alarm reappears, return to Step 1.
4
Press
MANUAL
.
5
Set the collect pump flow rate to 0 mL/min.
6
Set the centrifuge speed to 0 rpm.
7
Press
CHANGE MODE
, then press
4
to select “Rinseback.”
8
Follow the normal procedures to complete Rinseback and disconnect the donor/patient.
Note:
Alarm messages clear for 30 seconds. If alarm messages return while you perform the above steps,
remove the alarm messages and return to the Run screen by pressing
CLEAR
. If multiple alarms occur, it may be
necessary to press
CLEAR
several times.
Note:
If the multi-lumen tubing is twisted around the centrifuge arm, ensure the centrifuge arm is not bent.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...