Helpful Hints and Recovery Procedures
11-8
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Performing Manual Rinseback
Use a flat-blade screwdriver to perform this procedure.
If you decide to manually return the blood to the donor/patient,
perform the following:
1
Turn off the Spectra system’s power switch. Seal and remove the
collection bag(s).
2
Dual-Needle procedures:
Close the white pinch clamp on the
access needle line, then disconnect the access needle.
– or –
Single-Needle procedures:
Clamp the access line at the access
manifold.
3
Use the screwdriver to open the waste valve. This relieves the
pressure from the centrifuge.
4
Turn the valves to the positions shown in Table 11-2. Place a
screwdriver in the slot on the top of the valve or use your fingers to
turn the valve.
For information about valve positions, see “The VALVE Key” on
page 5-14. In particular, see “Waste Divert Valve” on page 5-15,
“Collect/Replace Valve” on page 5-16, and “Plasma/RBC Valve” on
page 5-16.
5
Remove the tubing from the return line valve.
6
Completely open the access saline line roller clamp.
7
Turn the inlet pump counterclockwise until you see that the RBC
line does not contain RBCs. Rotate the inlet pump by turning the
knob on the top of the pump.
Use approximately 200 mL of saline from the container.
8
Close the RBC valve.
Warning:
To avoid accidental removal of additional blood
Table 11-2:
Valve positions during manual Rinseback
Valve
Position
Waste Divert
Closed
Return Line
Open
RBC/Plasma Line
Open
Plasma/RBC
Return
Collect/Replace
Collect
Warning:
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...