Helpful Hints and Recovery Procedures
11-6
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Speeding or Slowing Rinseback
If you want to perform Rinseback mode faster or slower than the current
pump flow rates, perform the following:
1
After clamping and disconnecting the collection bags, press
CLEAR
to continue Rinseback mode.
2
When Rinseback enters the state in which you want to change the
speed, press
CLEAR
. The current pump flow rates appear.
3
Press the key for the pump flow rate you want to change.
4
Enter a new pump flow rate.
5
Press
ENTER
.
Determining Net Additional Saline Returned
to the Donor or Patient
See “Net Additional Saline Returned to the Donor or Patient” on
page 9-6.
Calculating Product Bag Tare Weights
For ELP, LRS, LRST, and WBC Procedures
To determine the tare weights of the product bags and applicable
attachments, perform the following:
1
Seal an empty platelet or WBC collection bag according to your
standard operating procedures.
2
Weigh the bag to establish the tare weight for this bag
configuration. Ensure you include all labels and tags normally in
place when you weigh product bags.
3
Subtract this tare weight from the weight of your product bag to
determine the weight of your product.
Calculating Spectra System Fluid Volumes
See Chapter 9,
.
Note:
During the recirculation phase of Rinseback, the
return line valve is closed. DO NOT adjust any pump flow
rate during the recirculation phase.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...