LED Program Board Overview
COBE 2991 Cell Processor
•
Essentials Guide
4-3
LED Progra
mming
4
Cycle
The arrow keys (located below the CYCLE LED) are used to select each cycle,
allowing parameters (TIME, R.C.O., STOP SELECT, and VALVE SELECT) to be
programmed, viewed, or modified. Up to seven cycles can be programmed for each
protocol. During processing in the Automatic mode, the cycle display indicates
which cycle the COBE 2991 is currently executing.
Spin Timer
The Spin Timer is used to control the duration of the Start/Spin operation for each
cycle. The spin time is set by pressing the corresponding arrow keys (located below
the TIME LED). The time range is from 00.0 to 99.9 minutes. The minimum
increment of the Spin Timer is 0.1 (1/10) of a minute, or six seconds.
Note:
During processing in the Automatic mode, elapsed spin time can be displayed by
pressing and holding the ELAPSED TIME key. Remaining spin time is displayed
again when the key is released. The remaining time can be increased or decreased
during the spin time countdown. When a spin operation is completed, the display
shows the total time elapsed during that Start/Spin operation.
R.C.O.
Select the R.C.O. (red cell override) option for a particular cycle by pressing the
R.C.O. key.
RCO is the delay interval between the time that RBCs are detected by the RCD
and the time the Supernatant-Out operation is halted. The RCO option is used to
discard a volume of unwanted RBCs following cell detection based on the SUPER
OUT RATE and RCO timer setting. See “RED CELL OVERRIDE (in Seconds)”
on page 3-11 for information on the RCO timer.
Summary of Contents for COBE 2991
Page 1: ......
Page 18: ...This page intentionally left blank ...
Page 19: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Page 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Page 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Page 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Page 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Page 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Page 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...