3-2
COBE 2991 Cell Processor
•
Essentials Guide
Operator Controls
Main Control Panel
Except where noted, the following information applies to both LED and pin-diode
program board versions of the COBE 2991.
CENTRIFUGE SPEED (in rpm)
The CENTRIFUGE SPEED control sets the speed of the centrifuge during the
Start/Spin and Supernatant-Out operations. The control is calibrated in rpm
(revolutions per minute) with an available range of 0 to 3000 rpm.
SUPER OUT RATE (in mL/min)
The SUPER OUT RATE control sets the rate at which fluids are expressed from the
cell processing bag during the Supernatant-Out operation. The control is calibrated
in milliliters per minute (mL/min) with an available range of 0 to 450 mL/min.
MINIMUM AGITATE TIME (in seconds)
The MINIMUM AGITATE TIME control sets the minimum duration of the
Agitate/Wash-In operation during processing in the Automatic mode. The control
is calibrated in eight distinct values, from 30 to 100 seconds, in 10-second
increments. The COBE 2991 continues the Agitate/Wash-In operation until the
preselected minimum agitate time has elapsed or the hydraulic pump returns to its
home (starting) position, whichever is longer.
SUPER OUT VOLUME (in mL)
The SUPER OUT VOLUME control sets the volume of supernatant that is
expressed during the Supernatant-Out operation. The control is calibrated in
milliliters (mL) with an available range of 40 to 600 mL. When the volume setting
is reached during the Supernatant-Out operation, the following occurs:
•
Automatic mode: the COBE 2991 advances to the Agitate/Wash-In operation,
and the hydraulic pump stops and reverses, returning to its home (starting)
position.
Note:
For the
specifications of the
above controls,
see “Specifications” on
page 9-1.
•
Manual mode: the hydraulic pump stops, the selected valve remains open, and
the centrifuge continues to spin.
Summary of Contents for COBE 2991
Page 1: ......
Page 18: ...This page intentionally left blank ...
Page 19: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Page 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Page 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Page 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Page 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Page 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Page 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...